Reported Jan 18, 2017 · Terminated
SafeStep¿ Huber Needle Set with Y-Injection Site. Product Code: LH-0029YN. Packaged as single units in thermoformed…
Bard Access Systems
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Bard Access Systems · 605 N 5600 W · Salt Lake City, UT · 84116-3738
Is this the product you have?
- SafeStep¿ Huber Needle Set with Y-Injection Site
- Product CodeLH-0029YN. Packaged as single units in thermoformed trays. There are 25 SafeStep¿ trays per case and 100 per carton
- Product UsageIntended for the administration of fluids and drugs, or blood sampling through surgically implanted vascular ports
- Lot codes and dates
- Lot NumberASZKS076
- Amount recalled
- 8000 units
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
Why it was recalled
Bard Access Systems announces a voluntary field action for the SafeStep¿ Huber Needle Set with Y-Injection Site and MiniLoc¿ Safety Infusion Set Winged without Y-Injection Site devices because sets may contain one or more units where the needle length and gage differs from the length and gage described on the label.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new packaged foods recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.