Reported Feb 26, 2014 · Terminated
Item codes starting with 001930 LAG SCR 12.7D X 55MM SHOR up to 155 MM, LAG SCR 12.7D X 55 MM SHO up to 100MM, LAG SCR…
Zimmer, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Zimmer, Inc. · 345 E Main St · Warsaw, IN · 46580-2746
Is this the product you have?
- Item codes starting with 001930 LAG SCR 12.7D X 55MM SHOR up to 155 MM, LAG SCR 12.7D X 55 MM SHO up to 100MM, LAG SCR 12.7D X 55MM STAN up to 155 MM, LAG SCR 15.8D X 55MM LARG up to 155 MM
- Lot codes and dates
1,478 of 1,478 codes
6148434962124275621571796107720961337135614843506055561160747379608748336114285461142854R6032555260293083608796836128433061286943600855276043602560513856608796846128433161286947601451676018685560210921608159406087968560152929602930846048314460523344604831456052334560483145R6044075660483146605233466114285660483147605233476053682060483148605233506048314960523353604831506052335560536821604831516048315260523360604831536052336460483154605233656106775560483155605233696013250460165597
- Amount recalled
- 7,044,680 total
This recall covers 44 products
The agency filed one notice per product. You are reading one of them; the other 43 are below.
31 more are in the archive.
Where it was sold
- Nationwide
Why it was recalled
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new packaged foods recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.