Reported Jan 29, 2025 · Ongoing
Brand Name: VITROS Product Name: VITROS XT 7600 Integrated System Model/Catalog Number: 6844461 (New), 6272222…
Ortho-Clinical Diagnostics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Ortho-Clinical Diagnostics, INc. · 513 Technology Blvd · Rochester, NY · 14626-3601
Is this the product you have?
- Brand NameVITROS
- Product NameVITROS XT 7600
- Integrated System Model/Catalog Number
6844461 (New)6272222 (Certified)6904003 (China)
- Software Versionversion 3.8.0
- or higher Product DescriptionIn vitro diagnostic chemistry analyzer
- FDA 3 letter product codeJJE
- ComponentNo, not a component
- Lot codes and dates
- Lot CodeVITROS
- XT 7600 Integrated System Product Code6844461 (New)
- UDI10758750012343 VITROS
- XT 7600 System (Certified) Product Code6272222
- UDI10758750012343
- Lot CodeVITROS
- XT 7600 Integrated System Product Code6904003 (China)
- UDI10758750035809
- Range of installed serial numbers76000001-76002199
- Manufacture date range
July 2018 to present Distribution date rangeJuly 2018 to present Total
- systems currently installed in the US994 Total systems
- currently installed outside the US899
- Note1The catalogue number 6844461 is used to denote the VITROS XT 7600 Integrated System. Catalogue number 6272222 is used to denote VITROS XT 7600 System Certified. A VITROS XT 7600 Integrated System maintains its original serial number through the certification process. Therefore, the manufacturing and distribution information for the two catalogue numbers will be combined
- Note2When manufactured the system s serial numbers are sequential, but not all analyzers are released for distribution and therefore not installed. The range of serial numbers provided includes the identified minimum and maximum in serial numbers associated with all currently installed systems. Therefore, the amount of each system manufactured may not equal the amount of each system installed. Additionally, not all serial numbers listed are associated with an active installed analyzer in use today
- Note3The value given for "Product Quantity Distributed" is the total number of active installed systems that are in use today. This number is known to fluctuate as individual systems are installed and/or decommissioned
- Amount recalled
- 1,893 units (994 US, 899 OUS)
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Where it was sold
- Nationwide
Why it was recalled
The VITROS XT 7600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired products, which would increase the risk of no results or erroneous results from the instrument.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Ortho-Clinical Diagnostics, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.