Reported Jan 7, 2026 · Ongoing

BVI Leos Laser and Endoscopy System. Model Number: OME6000U.

Beaver-Visitec International, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Beaver-Visitec International, Inc. · 500 Totten Pond Rd · Waltham, MA · 02451-1916

Is this the product you have?

  • BVI Leos Laser and Endoscopy System
  • Model NumberOME6000U
Lot codes and dates
  • Model NumberOME6000U
  • UDI(01)00886158037576(11)250513(21)
  • Serial Numbers
    • 77UBV500003
    • 77UBV500004
    • 77UBV500005
    • 77UBV500006
    • 77UBV500010
Amount recalled
5 units

Where it was sold

Why it was recalled

Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Beaver-Visitec International, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 7:33 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.