Reported Mar 27, 2019 · Terminated
smith&nephew SMF(TM) STANDARD OFFSET STIKTITE(TM) COATED STEM 12/14 TAPER -1 SIZE, STEM, REF 713552510
Smith & Nephew, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Smith & Nephew, Inc. · 1450 E Brooks Rd · Memphis, TN · 38116-1804
Is this the product you have?
- smith&nephew SMF(TM) STANDARD OFFSET STIKTITE(TM) COATED STEM 12/14 TAPER -1 SIZE, STEM, REF 713552510
- Lot codes and dates
- Batch Numbers
11AM15163U11AM15184T11CM01097U15MM0646616EM1301816EM13035 & 18CM08027
- Batch Numbers
- Amount recalled
- 58 units
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
Complaints were received indicating that the outer label incorrectly identified the product as size "1" instead of "-1". The product contained within the package is size "-1"