Reported Mar 27, 2019 · Terminated
CTL Medical Hex Driver, 1/4" Square QC, 4.0 Ratcheting Ref 015.7050
CTL Medical Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: CTL Medical Corporation · 4550 Excel Pkwy Ste 300 · Addison, TX · 75001-5714
Is this the product you have?
CTL Medical Hex Driver1/4" Square QC4.0 Ratcheting Ref 015.7050
- Lot codes and dates
Lot 7EQAB-D7ATAB-D
- Amount recalled
- 133 total devices
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Why it was recalled
the anti-toggle ring of the hex driver may separate from the driver shaft. If this occurs during surgery, the anti-toggle ring can be pressed into the pedicle screw head and cause separation of the screw from the housing upon tightening. if the anti-toggle ring remains in the pedicle crew head and can interfere with final tightening of the pedicle screw construct. this can lead to failure of the set crew and relies of the rod from the construct.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new CTL Medical Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.