Reported Feb 14, 2024 · Ongoing
BD BBL Sensi Disc Meropenem- 10 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231703…
Becton Dickinson & Co.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Becton Dickinson & Co. · 7 Loveton Cir · Sparks, MD · 21152-9212
Is this the product you have?
- BD BBL Sensi Disc Meropenem- 10 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231703 and 231704¿
- Lot codes and dates
- Catalog No. 231703 UDI-DI 00382902317036 Lots 2147649 9065664; Catalog No. 231704¿ UDI-DI 30382902317044 Lots 2213585 3061281 3090652 3213825 2187975 2213583
- Amount recalled
- 2,363,168 total units
This recall covers 30 products
The agency filed one notice per product. You are reading one of them; the other 29 are below.
17 more are in the archive.
Where it was sold
- Nationwide
Why it was recalled
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Becton Dickinson & Co. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.