Reported Feb 5, 2025 · Ongoing
HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 10mmx10mm UPN: M00553540
Boston Scientific Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Boston Scientific Corporation · 100 Boston Scientific Way · Marlborough, MA · 01752-1234
Is this the product you have?
- HOT AXIOS Stent and Electrocautery-
- Enhanced Delivery System 10mmx10mm UPNM00553540
- Lot codes and dates
- GTIN08714729904564
- Lot Numbers
33169261331692633320912733243075332430773324385633255964334883283378703133787032338533953388222133882222338996563390928033979053339790553402724534027246340355233403552434052776340527783416183134161832341618333416183434190943341909443419094534192326341923273420191734203537342035393420479534223677342236793422639934226540342434273424342834375175343818033458540834731098
- Amount recalled
- 519 units (OUS)
This recall covers 14 products
The agency filed one notice per product. You are reading one of them; the other 13 are below.
1 more are in the archive.
Where it was sold
- Nationwide
- PR
Why it was recalled
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Boston Scientific Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.