Discovery XR656 HD. X-Ray imaging system.
GE Medical Systems, LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: GE Medical Systems, LLC · 3000 N Grandview Blvd · Waukesha, WI · 53188-1615
Is this the product you have?
- Discovery XR656 HD. X-Ray imaging system.
- Lot codes and dates
- Definium Tempo (GTIN00195278118356)
- Definium Tempo Pro (GTIN00195278070265)
- Amount recalled
- 3149 (1697 US; 1452 OUS)
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
- Nationwide
- PR
Why it was recalled
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.