Reported Feb 11, 2015 · Terminated
ABL90 FLEX analyzer; Model number 393-090. A portable, automated analyzer that measures pH, blood gases, electrolytes,…
Radiometer America Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Radiometer America Inc · 810 Sharon Dr · Westlake, OH · 44145-1521
Is this the product you have?
- ABL90 FLEX analyzer; Model number 393-090. A portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood.
- Lot codes and dates
- Model number 393-090 (instrument). The affected analyzers are serial numbers 393-090R0027N001 to present.
- Amount recalled
- 5002 units
Where it was sold
Why it was recalled
The ABL90 analyzer does not always use the most recent calibration data to calculate patient results. This can in some cases lead to a biased patient result.
This product could cause temporary or reversible health problems — stop using it.