Reported Apr 3, 2019 · Terminated
BD Synapsys Laboratory Solutions Catalog Number:444150 Product Usage: BD Synapsys is a laboratory software solution…
Becton Dickinson & Co.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Becton Dickinson & Co. · 7 Loveton Cir · Sparks, MD · 21152-9212
Is this the product you have?
- BD Synapsys
- Laboratory Solutions Catalog Number444150
- Product UsageBD Synapsys is a laboratory software solution providing data management and workflow management functionality across clinical diagnostic activities. The software integrates operational and clinical data points from laboratory hardware, with patient data received from external software systems allowing for the aggregation, evaluation, dissemination of knowledge using all data, anywhere, anytime. Currently Release 2 of BD Synapsys can connect to both Lab Automation (Kiestra) and BACTEC instruments
- Lot codes and dates
- BD Synapsys version 2.1
- UDI00382904441500
- Amount recalled
- 5 sites
Where it was sold
Why it was recalled
BD Synapsys version 2.1 allowed the potential for a test order to be associated to an incorrect culture record.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Becton Dickinson & Co. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.