Reported Jan 21, 2026 · Ongoing
Product Name: Allura Xper FD10C; Model number: 722001;
Philips Medical Systems Nederland B.v.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Veenpluis 4-6 · Best · N/A
Is this the product you have?
- Product NameAllura Xper FD10C
- Model number722001
- Lot codes and dates
- Model Number722001
- System Product NameAllura Xper FD10C
- UDI-DIN/A
- Serial Numbers
1661811532742962672062471612822132841031991299529930235221348283276253110078229232159177219217205389101295305
- Amount recalled
- 38 units
This recall covers 18 products
The agency filed one notice per product. You are reading one of them; the other 17 are below.
5 more are in the archive.
Where it was sold
- Nationwide
Why it was recalled
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips Medical Systems Nederland B.v. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.