Reported Feb 1, 2017 · Terminated
Cobas b 123 POC system The cobas b 123 POC system is a fully automated POC system for whole blood in vitro measurement…
Roche Diagnostics Operations, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Roche Diagnostics Operations, Inc. · 9115 Hague Rd · Indianapolis, IN · 46256-1025
Is this the product you have?
- Cobas b 123 POC system The cobas b 123 POC system is a fully automated POC system for whole blood in vitro measurement of pH, blood gases (B3G), electrolytes Nat, K+, iCaWt (ISE), hematocrit (THct), metabolites (Glu, Lac), total hemoglobin (tHb), hemoglobin derivatives (O2Hb, HHb, COT~b, MetHb), and oxygen saturation (SO2). In addition, the cobas b 123 POC system calculates derived parameters. It is dedicated for use in a Point-of-Care environment and laboratory. The integrated AutoQC module and the oximeter module are available as an option
- Lot codes and dates
- Modelscobas b 123 <2>
- POC system05122252001 cobas b 123 < 4 >
- POC system05122287001
- Serial numbers
1107422742550230223042309125324202337214723892418242517492462233823391850232825001374230113492400234423501860194120062128
- Amount recalled
- 30
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
under specific settings, an issue may occur during simultaneous Sensor Cartridge and Fluid Pack change on the cobas b 123 <2> POC system and cobas b 123 <4> POC system. The issue occurs when the software function [AutoQC as follow-up] is configured to run all three levels of AutoQC only after a Fluid Pack change, but not after a Sensor Cartridge change. When both are changed simultaneously, starting with the Sensor Cartridge and followed by the Fluid Pack, the analyzer carries out only the follow-up actions associated with the Sensor Cartridge change after completing the change workflow. As a result, no follow-up AutoQC is performed and the three expected AutoQC measurements for the Fluid Pack change are not carried out. Without running quality control, there is a remote possibility that system issues would not be detected and wrong results would not be excluded on all parameters: pH, PO2, PCO2, Na+, K+, Ca++, Cl-, Glu, Lac, Hct, SO2, O2Hb, COHb, MetHb, HHb, and Bili.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Roche Diagnostics Operations, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.