Reported Feb 21, 2024 · Ongoing
MobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and 712007
Sedecal Sa
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: SEDECAL SA · PELAYA, 9 POL. · Algete, N/A · N/A
Is this the product you have?
MobileDiagnost wDRModel numbers 712001712002712004712006and 712007
- Lot codes and dates
5 of 5 products
Products covered by this recall Model / number Barcodes Product as published 71200100884838028586UDI-DI 00884838028586, all serial numbers 71200200884838082373UDI-DI 00884838082373, all serial numbers 71200408436046002173UDI-DI 08436046002173, all serial numbers 71200600884838091504UDI-DI 00884838091504, serial numbers: SN20410556 and SN21410184 71200708436046002357UDI-DI)08436046002357, serial numbers: 26, 27, 19410038, 19410044, 19410046, 21410261, 21410266, 21410280, 21410288, 21410290, 21410291, 21410293, 21410294, 21410339, 21410340, 21410341, 21410342, 21410345, 22410022, 22410023, SN0410157, SN1410049, SN19410034, SN19410035, SN19410036, SN19410037, SN19410071, SN19410092, SN19410093, SN19410094, SN19410097, SN19410098, SN19410099, SN19410100, SN19410101, SN19410102, SN19410115, SN19410116, SN19410123, SN19410127, SN19410129, SN19410135, SN19410136, SN19410137, SN20410029, SN20410030, SN20410032, SN20410033, SN20410034, SN20410044, SN20410045, SN20410049, SN20410060, SN20410063, SN20410064, SN20410065, SN20410066, SN20410071, SN20410072, SN20410073, SN20410074, SN20410081, SN20410082, SN20410083, SN20410084, SN20410093, SN20410094, SN20410095, SN20410096, SN20410097, SN20410115, SN20410116, SN20410117, SN20410118, SN20410119, SN20410120, SN20410154, SN20410171, SN20410172, SN20410202, SN20410218, SN20410219, SN20410242, SN20410260, SN20410261, SN20410279, SN20410280, SN20410288, SN20410289, SN20410290, SN20410291, SN20410298, SN20410299, SN20410300, SN20410313, SN20410321, SN20410334, SN20410360, SN20410367, SN20410373, SN20410395, SN20410396, SN20410428, SN20410472, SN20410473, SN20410482, SN20410555, SN20410556, SN20410557, SN20410566, SN20410567, SN20410570, SN21410017, SN21410034, SN21410035, SN21410048, SN21410060, SN21410061, SN21410062, SN21410081, SN21410093, SN21410094, SN21410094, SN21410099, SN21410103, SN21410104, SN21410105, SN21410117, SN21410118, SN21410119, SN21410130, SN21410131, SN21410132, SN21410133, SN21410149, SN21410150, SN21410153, SN21410154, SN21410155, SN21410169, SN21410170, SN21410171, SN21410172, SN21410187, SN21410188, SN21410189, SN21410190, SN21410200, SN21410201, SN21410203, SN21410204, SN21410205, SN21410206, SN21410207, SN21410208, SN21410212, SN21410213, SN21410214, SN21410232, SNZY20410009, SNZY20410011, SNZY20410014, and SNZY204110015. - Amount recalled
- 851 devices
Where it was sold
- Nationwide
Why it was recalled
While cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button. This would result in the system moving on its own in a very slow speed forward circular motion.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new infant & baby recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.