Reported Jan 21, 2026 · Ongoing
Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G
Medline Industries, LP
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Medline Industries, LP · 3 Lakes Dr · Northfield, IL · 60093-2753
Is this the product you have?
- Medline medical
- procedure convenience kits labeled asKIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G
- Lot codes and dates
- UDI/DI 10198459156991 (EA) 40198459156992 (CS), Lot Number 25EBM426
- Amount recalled
- 12 units
This recall covers 18 products
The agency filed one notice per product. You are reading one of them; the other 17 are below.
5 more are in the archive.
Where it was sold
- Nationwide
Why it was recalled
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Medline Industries, LP recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.