Reported Apr 24, 2019 · Terminated
icumedical ChemoLock Closed Vial Spike w/Skirt, (a) REF CL-80S-4 (4 units) (b) REF CL-80S-5 (5 units). Used in IV…
ICU Medical, Inc.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: ICU Medical, Inc. · 951 Calle Amanecer · San Clemente, CA · 92673-6212
Is this the product you have?
- icumedical ChemoLock Closed Vial Spike w/Skirt, (a) REF CL-80S-4 (4 units) (b) REF CL-80S-5 (5 units). Used in IV administration sets. The ChemoLock Closed System Drug Transfer Device prevents the transfer of environmental contaminants, including bacterial and airborne contaminants into the system, and the escape of drug or vapor concentrations outside the system. The ChemoLock is needlefree and cannot be deactivated, which will passively aid in preventing needlestick injuries and the exposure to cytotoxic medications for healthcare personnel.
- Lot codes and dates
4 of 4 products
Products covered by this recall Model / number Product as published CL-80S-4— — Lot Numbers: 3820520, 3820521, 3844785 (b) CL-80S-5— — Lot Numbers: 3820516, 3820523, 3839238, 3839239, 3844781, 3844782, 3853998, 3853999, 3866769, 3866770, 3872090, 3910645, 3932908, 3932910, 3946770, 3951236 - Amount recalled
- 23800 devices
This recall covers 7 products
The agency filed one notice per product. You are reading one of them; the other 6 are below.
Where it was sold
- Nationwide
Why it was recalled
Stop using this product now
There is a potential for certain lots of ChemoLock and ChemoClave Vial Spikes to have plastic burr particulate originating from the protective cap used in the assembly of the device.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new ICU Medical, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.