Reported May 25, 2022 · Terminated
CK MB Calibrator, Catalogue Number CK2393
Randox Laboratories Ltd.
Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Randox Laboratories Ltd. · 55 Diamond Road · Crumlin (North), N/A · N/A
Is this the product you have?
- CK MB Calibrator, Catalogue Number CK2393
- Lot codes and dates
- GTIN 05055273201451 Lot 4410CK
- Amount recalled
- 1,392
Where it was sold
Why it was recalled
Randox have realigned CK-MB Calibrator, CK2393, lot number 4410CK. Calibrating with the new value will cause a positive shift of approximately 6% with both Quality Controls and Patient Samples.