Reported Feb 19, 2025 · Ongoing

The MSOT Acuity Echo is a Class 4 medical laser device. The MSOT Acuity is a mobile assembly of electrically-powered…

Ithera Medical GmbH

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Ithera Medical Gmbh · Zielstattstr. 13 · Munich · N/A

Is this the product you have?

  • The MSOT Acuity Echo is a Class 4 medical laser device. The MSOT Acuity is a mobile assembly of electrically-powered components intended for imaging and analysis of soft-tissue and vasculature using combined optoacoustic (photoacoustic) and ultrasound (US) imaging.
Lot codes and dates
  • Model numbers MSOT Acuity Echo (Research System) MSOT Acuity
  • Echo CE MSOT Acuity Echo CE2 UDI's(01)04262380070004 (01)04262380070004(21)22102 (01)04262380070004(21)22203 (01)04262380070004(21)21901 (01)04262380070004(21)22101 (01)04262380070004(21)22104 (01)04262380070004(21)22105 (01)04262380070004(21)22106 (01)04262380070004(21)22107 (01)04262380070004(21)22201 (01)04262380070004(21)22202 (01)04262380070004(21)22204 (01)04262380070004(21)22205 (01)04262380070004(21)22206 (01)04262380070004(21)22301 (01)04262380070004(21)22401
Amount recalled
23 units (6 units US)

Where it was sold

  • Nationwide

Why it was recalled

The MSOT Acuity Echo does not include an adequate instruction for safe operation of the device. The previous version of the user manuals contained a wrong value for the Nominal Ocular Hazard Distance (NOHD).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Ithera Medical GmbH recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 7:42 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.