Reported Feb 1, 2017 · Terminated
ProSpore 4 mL Ampoule Biological indicator for steam sterilization, contains Geobacillus stearothermophilus, Catalog…
Mesa Laboratories Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Mesa Laboratories Inc · PO Box 27261 · Omaha, NE · 68127-3623
Is this the product you have?
- ProSpore 4 mL Ampoule Biological indicator for steam sterilization, contains Geobacillus stearothermophilus
- Catalog Number
PS-6-50contains 50 Ampoules/Box
- Product UsageBiological indicator for steam sterilization at 121 deg C intended for use by a health care provider to accompany products being sterilized through steam sterilization procedure and to monitor adequacy of sterilization
- Lot codes and dates
Lot # 637exp. date 12/2016
- Amount recalled
- 328/ 50-count/ boxes; 16,400 individual units
Where it was sold
Why it was recalled
Product resistance has fallen outside of label claim within expiry.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Mesa Laboratories Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.