Reported Feb 17, 2021 · Ongoing

Infinity Acute Care System (IACS) Monitoring Solution: The IACS is intended for multi-parameter, physiologic patient…

Draegar Medical Systems, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Draegar Medical Systems, Inc. · 6 Tech Dr · Andover, MA · 01810-2434

Is this the product you have?

  • Infinity Acute
  • Care System (IACS) Monitoring SolutionThe IACS is intended for multi-parameter, physiologic patient monitoring of adult, pediatric and neonatal patients in environments where patient care is provided by trained healthcare professionals
Lot codes and dates
    • Model Number - MS25510
    • MS25520
    • MS26372 Revisions 24 to 29
Amount recalled
19802 systems (Domestic: 3412, Foreign 16,390)

Where it was sold

Why it was recalled

The Infinity M540 patient monitor may randomly reboot due to an error to correctly transmit and read the header data of files in the memory of the device. Under this situation, the device will try to reboot to mitigate the error. The device will be available again for use within 30 seconds. If this error continues and the M540 reboots three times in a time span of 10 minutes, it will enter a fail-state. A fail-state will annunciate itself with a continuous sound to alert the user. The M540 will reset to factory default and the user can manually configure and readmit the patient to continue patient monitoring.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Draegar Medical Systems, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:55 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.