Reported Apr 24, 2013 · Terminated
REF 32-422760 Oxford Uni Knee System,Toffee Hammer, non-sterile, found in REF 32-422763 Oxford Mlcroplasty Knee System…
Biomet U.K., Ltd.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Biomet U.K., Ltd. · Waterton Industrial Estate · Bridgend South Wales, N/A · N/A
Is this the product you have?
- REF 32-422760 Oxford Uni Knee System,Toffee Hammer, non-sterile, found in REF 32-422763 Oxford Mlcroplasty Knee System Tibial instrument Tray small surgical hammer supplied with the Oxford Partial Knee Microplasty Instrument set, used for impacting implants.
- Lot codes and dates
- 32-422760 ( sold as part of instrument set 32-422763 ) ZB100901,ZB110201,ZB110301,ZB110302,ZB110303, ZB110304,ZB110601,ZB110602,ZB110701,ZB111201, ZB111202,ZB120201,ZB120202,ZB120701,ZB120801, ZB120802,ZB121102,ZB121103
- Amount recalled
- 430
Where it was sold
- Nationwide
Why it was recalled
Investigation of complaints found excessive use of the instrument could result in cracking on the weld between the head and the handle, which could lead to part of the head becoming detached from the handle.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Biomet U.K., Ltd. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.