Reported Feb 25, 2015 · Terminated
Cold Therapy Pad - cold water circulating pad, one unit per package. The pads are used to provide a patient with local…
Cincinnati Sub-Zero Products Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Cincinnati Sub-Zero Products Inc · 12011 Mosteller Rd · Cincinnati, OH · 45241-1528
Is this the product you have?
- Cold Therapy Pad - cold water circulating pad, one unit per package. The pads are used to provide a patient with local cold therapy by circulating chilled water through the pad that has been positioned on the patient.
- Lot codes and dates
- ModelCatalog #CT -99
- Part number50137
- Lot number(s)
1338013390 and 13418
- Amount recalled
- 1390
Where it was sold
Why it was recalled
Seals in the water path of the CT-99 cold therapy pad may separate during use, which could lead to a water leak.
This product could cause temporary or reversible health problems — stop using it.