Reported Feb 25, 2015 · Terminated

Cold Therapy Pad - cold water circulating pad, one unit per package. The pads are used to provide a patient with local…

Cincinnati Sub-Zero Products Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Cincinnati Sub-Zero Products Inc · 12011 Mosteller Rd · Cincinnati, OH · 45241-1528

Is this the product you have?

  • Cold Therapy Pad - cold water circulating pad, one unit per package. The pads are used to provide a patient with local cold therapy by circulating chilled water through the pad that has been positioned on the patient.
Lot codes and dates
  • ModelCatalog #CT -99
  • Part number50137
  • Lot number(s)
    • 13380
    • 13390 and 13418
Amount recalled
1390

Where it was sold

Why it was recalled

Seals in the water path of the CT-99 cold therapy pad may separate during use, which could lead to a water leak.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Cincinnati Sub-Zero Products Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 19, 2015 · Data as of Sep 19, 5:25 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.