HORIBA ABX PENTRA 400 I.S.E. Module, Model Numbers: P400ISE110EN02 and P400ISE110US02, All Serial Numbers, Software…
Horiba Instruments, Inc. dba Horiba Medical
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Horiba Instruments, Inc dba Horiba Medical · 34 Bunsen · Irvine, CA · 92618-4210
Is this the product you have?
- HORIBA ABX PENTRA 400 I.S.E. Module
- Model NumbersP400ISE110EN02 and P400ISE110US02, All Serial Numbers, Software Versions 5.0.7 and Below. Used to measure absorbance and ion electrodes
- Lot codes and dates
- Model NumbersP400ISE110EN02 and P400ISE110US02. All Serial Numbers. Versions 5.0.7 and below
- Amount recalled
- 905 units
Where it was sold
- Nationwide
Why it was recalled
HORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All other assays run on this analyzer do have Linearity Limit Flags to alert the user to values that fall below or above the validated linearity range for each assay.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Horiba Instruments, Inc. dba Horiba Medical recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.