Reported Mar 28, 2018 · Terminated
BD Vacutainer¿ Lithium Heparin Green Top Tube; Catalog Numbers: 1. 366664 (UDI: 50382903666640); 2. 367884 (UDI:…
Becton Dickinson & Company
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Becton Dickinson & Company · 1 Becton Dr · Franklin Lakes, NJ · 07417-1815
Is this the product you have?
6 of 6 products
| # | Barcodes | Product as published |
|---|---|---|
| 1 | 50382903666640 | 366664 (UDI: 50382903666640) |
| 2 | 50382903678841 | 367884 (UDI: 50382903678841) |
| 3 | 50382903678865 | 367886 (UDI: 50382903678865) |
| 4 | 50382903680400 | 368040 (UDI: 50382903680400) |
| 5 | 50382903666671 | 366667 (UDI: 50382903666671) |
| 6 | 50382903678803 | 367880 (UDI: 50382903678803) |
- Lot codes and dates
- All lots
- Catalog Number (Shelf Life)
1. 366664 (Requested)2. 367884 (Requested)3. 367886 (Requested)4. 368040 (Requested)5. 366667 (Requested)6. 367880 (Requested)
- Amount recalled
- 117,816,603
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
Why it was recalled
Stop using this product now
BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Becton Dickinson & Company recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.