Reported Mar 4, 2015 · Terminated

NAMIC Convenience Kit, UPN H749601314811, REF/Catalog No. 60131481, STERILE, Rx ONLY --- For single use only. ---…

Navilyst Medical, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Navilyst Medical, Inc · 10 Glens Falls Technical Park · Glens Falls, NY · 12801

Is this the product you have?

  • NAMIC Convenience Kit, UPN H749601314811, REF/Catalog No. 60131481, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588
  • Product UsageNAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems
Lot codes and dates
    • Lots 4245021
    • 4260368
    • 4268542
    • 4279333
    • 4309640
    • 4344841
    • 4380268
    • 4387481
    • 4390493
    • 4413788
    • 4436076
    • 4436659
    • 4447611
    • 4474955
    • 4491914
    • 4503486
    • 4505497
    • 4534914
    • 4549037
    • 4573406
    • 4576438
    • 4603643
    • 4708214
    • 4714177
    • 4716348
    • 4726245
    • 4729318
    • 4734982
    • 4750219
    • 4750566
    • 4760725
    • 4766447
    • 4779497
    • 4780532
    • 4790816
    • 4802163
    • 4811301
    • 4819730
    • 4822657
Amount recalled
411 units

This recall covers 60 products

The agency filed one notice per product. You are reading one of them; the other 59 are below.

47 more are in the archive.

Where it was sold

  • Nationwide

Why it was recalled

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Navilyst Medical, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Mar 10, 2017 · Data as of Sep 19, 5:40 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.