Reported Feb 15, 2017 · Terminated
VITROS Chemistry Products FS Calibrator 1, REF 680 1873 12 x 3 mL, Rx ONLY, IVD; used in conjunction with VITROS…
Ortho-Clinical Diagnostics
Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Ortho-Clinical Diagnostics · 1000 Lee Road · Rochester, NY · 14606
Is this the product you have?
- VITROS Chemistry Products FS Calibrator 1, REF 680 1873 12 x 3 mL, Rx ONLY, IVD; used in conjunction with VITROS Chemistry Products Calibrator Kits to calibrate VITROS 5,1 FS/4600 Chemistry and integrated Systems. --- Ortho-Clinical Diagnostics, Inc. Rochester, NY 14626 --- Ortho-Clinical Diagnostics Pencoed, CF35 5PZ
- UK ---- BackgroundVITROS assays must be calibrated prior to use in order to be used to produce assay results. The Assay Data Disk (ADD) contains calibrator assigned values/ calculation data for multiple VITROS Calibrators and Lots. The Data Release Version (DRV) number identifies the version of the ADD. Each time the ADD is updated with new ata, the DRV number increments higher. The DRV with the highest number contains the most current data. The VITROS 4600 and 5,1 FS Chemistry Systems and the VITROS 5600 Integrated System are the only VITROS Systems that run the assays that are associated with VITROS Chemistry Products FS Calibrator 1. VITROS FS Calibrator 1 is required to calibrate VITROS RF, dLDL, hsCRP, AMPH, BARB, BENZ, COCM, CP, ASO, THC, d%A1c assays. VITROS Chemistry Products FS Calibrator 1 is used in conjunction with VITROS Chemistry Products Calibrator Kits to calibrate the VITROS 5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. VITROS FS Calibrator 1 is used on VITROS Systems
- to calibrate the following assaysVITROS AMPH, ASO, BARB, BENZ, COCM, d%A1c, dLDL, hsCRP, PCP, RF and THC Reagents
- Lot codes and dates
- Lot 91461, Expiry Date 2018-09-19; Affected Assay Data Diskettes (ADD) Release Versions 5904 through 5910
- Amount recalled
- US: 160 kits, Foreign: 642 kits
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
- Nationwide
- PR
Why it was recalled
The company received a customer complaint for the inability to calibrate VITROS Chemistry Products dLDL Reagent when using VITROS FS Calibrator 1, Lot 91461 and ADD (DRV 5908). Assay Data Disk (ADD) DRV versions 5904, 5905, 5906, 5907, 5908, 5909, and 5910 are all affected by this issue.
This product is unlikely to cause harm, but it breaks a safety or labelling rule.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Ortho-Clinical Diagnostics recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.