Reported Mar 23, 2016 · Terminated

Transport Stretcher, and Head and Neck Surgery Stretcher Non-powered stretcher

Stryker Medical Division of Stryker Corporation

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Stryker Medical Division of Stryker Corporation · 3800 E Centre Ave · Portage, MI · 49002-5826

Is this the product you have?

  • Transport Stretcher, and Head and Neck Surgery Stretcher Non-powered stretcher
Lot codes and dates
  • The
  • hydraulic jack assemblies (Part numbers0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) used as service parts on Stretchers assembled from 8/28/15 - 8/29/15
Amount recalled
36

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

The hydraulic jack assemblies (PN: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) were manufactured using the incorrect oil. The oil used does not contain the same wear resistance additives as the specified oils and therefore its use increases the potential for wear which may cause a leak event or jack drift. Testing was done to evaluate the jacks with incorrect oil and leak events we

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Stryker Medical Division recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Sep 6, 2016 · Data as of Sep 19, 5:20 PM UTC