Reported Feb 26, 2025 · Ongoing

Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 12.0.0.932,…

Raysearch Laboratories Ab

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: RAYSEARCH LABORATORIES AB · Eugeniavagen 18c · Stockholm, N/A · N/A

Is this the product you have?

  • Brand NameRayStation- Intended as a Radiation Therapy Treatment
  • Planning System Model/Catalog Number
    • 12.0.0.932
    • 12.1.0.1221
    • 12.0.3.68
    • 12.0.4.12 and 12.3.0.119
  • Software Version
    • RayStation 11B and 11B Service Pack 1
    • 2
    • 3 and Service Pack Toshiba 1
Lot codes and dates
  • Lot Code
    • UDI: 0735000201042620211208
    • 0735000201049520220312
    • 0735000201050120220422
    • 0735000201060020220620
  • and 0735000201057020221222 GTIN
    • 07350002010426
    • 07350002010495
    • 07350002010501
    • 07350002010600
  • and 07350002010570 Serial Numbers
    • 12.0.0.932
    • 12.1.0.1221
    • 12.0.3.68
    • 12.0.4.12 and 12.3.0.119
  • Software Revisions
    • RayStation 11B and 11B Service Pack 1
    • 2
    • 3 and
  • Service Pack Toshiba 1 Expiration Date2028-03-20
Amount recalled
5

This recall covers 11 products

The agency filed one notice per product. You are reading one of them; the other 10 are below.

Where it was sold

  • Nationwide

Why it was recalled

Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Raysearch Laboratories Ab recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 7:35 PM UTC