Reported Mar 4, 2015 · Terminated

NAMIC Convenience Kit, LEFT HEART KIT, UPN H9656002072121, REF/Catalog No. 600207212, STERILE, Rx ONLY --- For single…

Navilyst Medical, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Navilyst Medical, Inc · 10 Glens Falls Technical Park · Glens Falls, NY · 12801

Is this the product you have?

  • NAMIC Convenience Kit, LEFT HEART KIT, UPN H9656002072121, REF/Catalog No. 600207212, STERILE, Rx ONLY --- For single use only. ---Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
Lot codes and dates
    • Lots 4652732
    • 4656169
    • 4657278
    • 4673716
    • 4694148
    • 4729301
    • 4748825
    • 4778748
    • 4790216
    • 4816281
    • 4817327
    • 4830491
Amount recalled
18 units

This recall covers 60 products

The agency filed one notice per product. You are reading one of them; the other 59 are below.

47 more are in the archive.

Where it was sold

  • Nationwide

Why it was recalled

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Navilyst Medical, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Mar 10, 2017 · Data as of Sep 19, 5:25 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.