Reported Mar 19, 2014 · Terminated

Mar Cor Purification CWP Reverse Osmosis System. Models 101, 102, 104, and 106.

Mar Cor Purification

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Mar Cor Purification · 14550 28th Ave N · Minneapolis, MN · 55447-4817

Is this the product you have?

  • Mar Cor Purification CWP Reverse Osmosis System. Models 101, 102, 104, and 106.
Lot codes and dates
  • serial numbers

    130 of 130 codes

    • 1289064
    • 1289069
    • 1289159
    • 1289259
    • 1289387
    • 1289650
    • 1289747
    • 1289826
    • 1289949
    • 1290088
    • 1290151
    • 1290333
    • 1290367
    • 1290480
    • 1290489
    • 1290845
    • 1290988
    • 1291042
    • 1291120
    • 1291233
    • 1291373
    • 1291378
    • 1291634
    • 1291777
    • 1291828
    • 1291933
    • 1292031
    • 1292114
    • 1292185
    • 1292185
    • 1292192
    • 1292391
    • 1292546
    • 1292597
    • 1292670
    • 1292674
    • 1292899
    • 1292960
    • 1293010
    • 1293039
    • 1293185
    • 1293190
    • 1293311
    • 1293379
    • 1293414
    • 1293687
    • 1293724
    • 1293771
    • 1293777
    • 1293953
    • 1293957
    • 1294093
    • 1294103
    • 1294171
    • 1294307
    • 1294362
    • 1294433
    • 1294516
    • 1294589
    • 1294727
Amount recalled
130

Where it was sold

  • Nationwide

Why it was recalled

There exists a potential situation with the crimp connection at the pump head assembly. The potential exists that the reverse osmosis pump head wiring could build resistance resulting in resistive heating of the surrounding components. If this occurs, the pump head assembly may experience melting of the plastic components and potentially create smoke/fire and shut down of the unit.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Mar Cor Purification recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 22, 2014 · Data as of Sep 19, 4:47 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.