Reported Jun 8, 2022 · Completed

DAVINCI | Energy SynchroSeal da Vinci X' da Vinci Xi', Single Use Instrument 8mm, REF 480440

Intuitive Surgical, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Intuitive Surgical, Inc. · 1266 Kifer Rd Bldg 101 · Sunnyvale, CA · 94086-5304

Is this the product you have?

  • DAVINCI | Energy SynchroSeal da Vinci X' da Vinci Xi', Single Use Instrument 8mm, REF 480440
Lot codes and dates
  • UDI00886874117309. All lots
Amount recalled
7,995 boxes (or 47,972 individual units ; 6 instruments/box)* The total amount will continue to grow until updated user documentation is distributed

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide
  • PR

Why it was recalled

Placing excessive tissue in the instrument jaws of the devices prior to sealing and transection can result in an insufficient seal.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Intuitive Surgical, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:39 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.