Reported Mar 15, 2023 · Ongoing

LATITUDE NXT Remote Patient Management System, LATITUDE NXT System Server Software Model 6460.

Boston Scientific Corporation

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Boston Scientific Corporation · 4100 Hamline Ave N · Saint Paul, MN · 55112-5700

Is this the product you have?

  • LATITUDE NXT Remote Patient Management System, LATITUDE NXT System Server Software Model 6460.
Lot codes and dates
  • UDI-DI00802526613876
Amount recalled
1,531 patient data records

Where it was sold

Why it was recalled

Under specific circumstances, the U.S. product registration system did not send up enablement requests to U.S. LATITUDE Customer Support during an 8-month interval in 2022. For these patients, the HeartLogic Index and any associated Yellow Alerts are currently not visible in LATITUDE as may have been intended by the healthcare provider..

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Boston Scientific Corporation recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:58 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.