Reported Mar 5, 2025 · Ongoing

DxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software.

Beckman Coulter Laboratory Systems (suzhou) Co., Ltd.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: BECKMAN COULTER LABORATORY SYSTEMS (SUZHOU) CO., LTD. · No. 11 Shengpu Street Zhensheng Road · Suzhou, N/A · N/A

Is this the product you have?

  • DxC 500 AU Clinical Chemistry Analyzer
  • REF
    • C63519
    • and C63520 with affected software
Lot codes and dates

2 of 2 products · 56 lot codes

Recalled lot codes, grouped by product number
REFUDI-DISerial Numbers
C6351914987666545058
  • 2024090170
C6352014987666545065
  • 2024070116
  • 2023070009
  • 2024010038
  • 2024010035
  • 2024050072
  • 2024040056
  • 2024030049
  • 2024020046
  • 2024070111
  • 2024030051
  • 2024030052
  • 2024010033
  • 2023090015
  • 2024060103
  • 2024060104
  • 2024060101
  • 2024080145
  • 2024020045
  • 2024020042
  • 2024020043
  • 2024020047
  • 2024060096
  • 2024040059
  • 2023100019
  • 2024030053
  • 2023070010
  • 2024070129
  • 2024070130
  • 2024020041
  • 2024090183
  • 2024010036
  • 2024010037
  • 2024070132
  • 2024070131
  • 2023100020
  • 2024050064
  • 2024050070
  • 2023090016
  • 2023110022
  • 2024070110
  • 2024060075
  • 2024060106
  • 2023110023
  • 2024010024
  • 2024010025
  • 2024050066
  • 2024010029
  • 2024060107
  • 2023080012
  • 2024040060
  • 2024050074
  • 2024040062
  • 2024010031
  • 2024030055
  • 2024080144
Amount recalled
56

Where it was sold

Why it was recalled

Beckman Coulter has identified an issue which prevents the DxC 500 AU instrument from providing the correct onboard stability (Open Expiration) dates for assays that were loaded prior to a software upgrade (e.g. 1.3 to 1.4, 1.4 to 1.4.1, etc.). As a result, assays that have expired may still be used for sample processing, leading to inaccurate or erroneous patient test results. The issue only happens on assays loaded before instruments doing upgrades and does not impact new installation. It does not impact the assays loaded after upgrade.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Beckman Coulter Laboratory Systems (suzhou)… recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:39 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.