Reported Mar 11, 2015 · Terminated

Focal Sim radiation therapy treatment planning system

Elekta, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Elekta, Inc. · 400 Perimeter Center Ter Ne · Atlanta, GA · 30346-1227

Is this the product you have?

  • Focal Sim radiation therapy treatment planning system
Lot codes and dates
  • Focal Release 4.80
Amount recalled
907

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

Incorrect CT to ED when using the Monaco Image Statistics Tool.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Elekta, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 7, 2017 · Data as of Sep 19, 4:47 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.