Reported May 8, 2019 · Terminated
CAROLINA(R) LIQUID CHEMISTRIES CORP. ETHA RGT. KIT, IVD, REF ST321
Carolina Liquid Chemistries
Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Carolina Liquid Chemistries · 313 Gallimore Dairy Rd · Greensboro, NC · 27409-9724
Is this the product you have?
- CAROLINA(R) LIQUID CHEMISTRIES CORP. ETHA RGT. KIT, IVD, REF ST321
- Lot codes and dates
- Lot 11687
- Amount recalled
- 20 kits
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
Supplier testing showed that the absorbance rate of the Ethyl Alcohol Assay did not meet internal quality control specifications. The defective lots will not function toward the end of shelf life.
This product is unlikely to cause harm, but it breaks a safety or labelling rule.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Carolina Liquid Chemistries recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.