Reported Apr 4, 2018 · Terminated

F-T3SYSTEM, T3 (Trumpf Table Top) System The device is a surgical table accessory system used to allow prone…

Allen Medical Systems

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Allen Medical Systems · 100 Discovery Way · Acton, MA · 01720-4481

Is this the product you have?

  • F-T3SYSTEM, T3 (Trumpf Table Top) System The device is a surgical table accessory system used to allow prone positioning on OR table tops during spinal surgery. The system consists of a table top frame, arm supports,
Lot codes and dates
  • Serial Numbers

    101 of 101 codes

    • A166677
    • A201568
    • A311424
    • A333604
    • A353171
    • A385408
    • A166678
    • A201569
    • A311425
    • A339446
    • A357145
    • A385409
    • A170724
    • A201570
    • A314187
    • A339447
    • A359864
    • A385410
    • A179337
    • A209110
    • A319402
    • A348001
    • A362219
    • A386775
    • A182027
    • A209111
    • A320583
    • A348174
    • A364982
    • A386776
    • A183578
    • A209112
    • A320584
    • A348175
    • A369558
    • A386777
    • A188535
    • A209113
    • A322480
    • A348176
    • A369884
    • A386778
    • A194789
    • A282259
    • A322597
    • A348177
    • A370262
    • A390698
    • A194790
    • A282260
    • A325663
    • A348178
    • A372387
    • A390699
    • A197337
    • A282537
    • A325664
    • A348179
    • A378593
    • A390700
Amount recalled
83

Where it was sold

  • Nationwide

Why it was recalled

The recalling firm has become aware that the rail section clamps used to attach the leg support of the device may fail during use, which could result in critical harm to the patient.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Allen Medical Systems recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 6, 2020 · Data as of Sep 19, 5:33 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.