Reported Mar 18, 2015 · Terminated

Carton: Richard-Allan Scientific CYTO-STAIN; REF 7501R; Quantity: 1x1 Gallon; Size: 1 Gallon/3.8L; LOT 313833; 11/2016;…

Richard-Allan Scientific Company

Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Richard-Allan Scientific Company · 4481 Campus Dr · Kalamazoo, MI · 49008-2590

Is this the product you have?

  • CartonRichard-Allan Scientific CYTO-STAIN; REF 7501R
  • Quantity1x1 Gallon
  • Size
    • 1 Gallon/3.8L
    • LOT 313833
    • 11/2016
  • BottleRichard-Allan Scientific CYTO-STAIN G; REF 7501G; 1 Gallon/3.8L; Thermo Scientific
Lot codes and dates
  • Lot313833
  • Expiration date11/2016
Amount recalled
132

Why it was recalled

The case was correctly labeled as REF 7501R, however the gallon bottle was incorrectly labeled as REF 7501G Cyto-Stain G. The bottle label has the correct lot number of 313833 and EXP date 11/2016. The gallon bottle contains the correct intended contents of 7501R: Cyto-Stain R Stain.

This product is unlikely to cause harm, but it breaks a safety or labelling rule.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Richard-Allan Scientific Company recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jan 12, 2016 · Data as of Sep 19, 4:49 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.