Enzyme linked immunosorbent assay, for herpes simplex virus, hsv-2
Siemens Healthcare Diagnostics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Healthcare Diagnostics, Inc · 333 Coney St · East Walpole, MA · 02032-1516
Is this the product you have?
- Enzyme linked immunosorbent assay, for herpes simplex virus, hsv-2
- Lot codes and dates
- ***Updated as of 3/25/2021*** Lot #: B33255, B33269, B33413. UDI #: (01)00630414016726(10)B33255(17)20210625, (01)00630414016726(10)B33269(17)20210629, (01)00630414016726(10)B33413(17)20210819. Updated as of 3/25/2021
- Amount recalled
- 390 kits total (Domestic 366 kits, Foreign 24 kits)
Where it was sold
Why it was recalled
Potential false positive HSV results. Siemens Healthcare Diagnostics, Inc. initiated a correction for ADVIA Centaur4P Herpes-2 lgG (HSV2) Onboard Stability and Calibration Frequency (Interval) Failures on the ADVIA Centaur XP System and ADVIA Centaur XPT System.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.