Reported Apr 6, 2016 · Terminated

Estradiol II Elecsys and cobas e analyzers REF 03000079 190; M 6.5mL, R1 8 mL, R2 8 mL, IVD, For USA: CONTENT M…

Roche Diagnostics Operations, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Roche Diagnostics Operations, Inc. · 9115 Hague Rd · Indianapolis, IN · 46256-1025

Is this the product you have?

5 of 5 products

Products covered by this recall
Model / numberProduct as published
03000079Estradiol II Elecsys and cobas e analyzers 190
M 6.5mL, R1 8 mL, R2 8 mL, IVD, For USA: CONTENT M Streptavidin-coated microparticles 0.72 mg/mL
R1 Biotinylated polyclonal anti-estradiol antibody (rabbit) 45 ng/mL
Mesterolone 130 ng/mL
R2 Estradiol derivative, labeled with ruthenium complex 2.75 ng/mL Product Usage: Immunoassay for the in vitro quantitative determination of estradiol in human serum and plasma. The electrochemiluminescence immunoassay ECLIA is intended for use on Elecsys and cobas e immunoassay analyzers.
Lot codes and dates
  • Part Number Elecsys Estradiol II03000079190; all lot numbers are affected
Amount recalled
102,312

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide
  • PR

Why it was recalled

Due to the risk of a recently identified cross reactivity, the Elecsys Estradiol assay should not be used when monitoring patient being treated with fulvestrant. Customers should be informed that fulvestrant will increase the apparent concentration of estradiol in women being treated with this drug. If treatment with fulvestrant has been altered or discontinued as a result of falsely elevated e

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Roche Diagnostics Operations, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jan 10, 2018 · Data as of Sep 19, 6:02 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.