Reported Mar 20, 2024 · Completed
LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004
Fresenius Kabi USA, LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Fresenius Kabi USA, LLC · 50 High St Ste 50 · North Andover, MA · 01845-2620
Is this the product you have?
- LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004
- Lot codes and dates
- UDI-DI 00811505030122 Software versions 5.0.1 and earlier.
- Amount recalled
- 3
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable (Fail-Stop) Pump Problem alarm. This may lead to delay of therapy, and patient may experience from no to minimal to serious harm. This issue was resolved in all fielded product in software version 5.2.1, which was installed in September 2022.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Fresenius Kabi USA, LLC recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.