Reported Mar 20, 2024 · Completed

LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004

Fresenius Kabi USA, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Fresenius Kabi USA, LLC · 50 High St Ste 50 · North Andover, MA · 01845-2620

Is this the product you have?

  • LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004
Lot codes and dates
  • UDI-DI 00811505030122 Software versions 5.0.1 and earlier.
Amount recalled
3

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable (Fail-Stop) Pump Problem alarm. This may lead to delay of therapy, and patient may experience from no to minimal to serious harm. This issue was resolved in all fielded product in software version 5.2.1, which was installed in September 2022.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Fresenius Kabi USA, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:42 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.