Reported Apr 11, 2018 · Terminated
DS Titanium Ligation Clip, used in laparoscopic surgery
Aesculap Implant Systems LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Aesculap Implant Systems LLC · 3773 Corporate Pkwy · Center Valley, PA · 18034-8217
Is this the product you have?
- DS Titanium Ligation Clip, used in laparoscopic surgery
- Lot codes and dates
- DS Clips - Product Numbers PL450SU DS SINGLE FIRE LIG.CLIPS PL471SU DS SINGLE FIRE LIG.CLIP L W/LATCH PL453SU DS SINGLE FIRE LIG.CLIP SM PL475SU DS SINGLE FIRE LIG.CLIP XL W/LATCH PL459SU DS SINGLE FIRE LIG.CLIP M PL462SU DS SINGLE FIRE LIG.CLIP ML PL465SU DS SINGLE FIRE LIG.CLIP ML W/LATCH PL468SU DS SINGLE FIRE LIG.CLIP L (all clips going back 5 years)
- Amount recalled
- 9621
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Why it was recalled
Added a contraindication to IFU: Do not use the DS clips for living donor nephrectomies. Added other contraindications. A publication from 2017 recommended Aesculap DS Clips for kidney transplantation with living donors. The information from this publication was not approved or validated by the firm.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Aesculap Implant Systems LLC recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.