Reported Mar 25, 2015 · Terminated
Crea A and B membranes, packaged 4 per box Used in conjunction with the ABL800 FLEX analyzers which are intended for:…
Radiometer America Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Radiometer America Inc · 810 Sharon Dr · Westlake, OH · 44145-1521
Is this the product you have?
- Crea A and B membranes, packaged 4 per box Used in conjunction with the ABL800
- FLEX analyzers which are intended forIn Vitro Testing of samples of whole blood for the parameters pCO2, cK+,cNa+, cCa2+,cCl-,cGlu, cLac, cCrea, ctBil, and co-oximetry parameters,(ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF), in vitro testing of samples of expired air for the parameters pO2 and pCO2, in vitro testing of pleura samples for the pH parameter
- Lot codes and dates
- Model Number 942-073; all lots which have not expired their product lifetime which
- would be lot 972-073R0465 and onwardAffected lot numbers
82 of 82 codes
942-073R0465942-073R0468942-073R0469942-073R0470942-073R0471942-073R0472942-073R0474942-073R0475942-073R0476942-073R0478942-073R0479942-073R0480942-073R0481942-073R0482942-073R0487942-073R0488942-073R0489942-073R0490942-073R0491942-073R0492942-073R0493942-073R0494942-073R0495942-073R0496942-073R0497942-073R0498942-073R0499942-073R0500942-073R0501942-073R0502942-073R0503942-073R0504942-073R0507942-073R0509942-073R0510942-073R0511942-073R0512942-073R0513942-073R0515942-073R0518942-073R0520942-073R0521942-073R0522942-073R0523942-073R0524942-073R0525942-073R0526942-073R0527942-073R0528942-073R0529942-073R0530942-073R0532942-073R0533942-073R0534942-073R0535942-073R0536942-073R0537942-073R0538942-073R0539942-073R0540
- And onwardsThe firm is still determining the scope of how many lot numbers are affected
- Amount recalled
- 10,824 boxes
Where it was sold
Why it was recalled
Negative drift on QC during in-use lifetime of CREA membranes can be observed.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Radiometer America Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.