Reported Mar 12, 2025 · Ongoing

TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP

TAS Medical Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: TAS Medical Inc · 9100 Conroy Windermere Rd Ste 200 · Windermere, FL · 34786-8431

Is this the product you have?

  • TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP
Lot codes and dates

3 of 3 products

Products covered by this recall
#BarcodesProduct as published
100197644217267T-4000/00197644217267/230403(2024/04/20)
200860011435921T-5000/00860011435921/240801(2025/08/28), 240901(2025/09/25)
30086001143591400197644927241T-LAP/00860011435914, 00197644927241/240702(2025/07/08), 230906(2024/11/29), 230905(2024/11/29)
Amount recalled
114

Where it was sold

Why it was recalled

Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new TAS Medical Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:45 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.