Reported Apr 11, 2018 · Terminated
Binding Site Optilite Beta-2-Microglobulin Urine Kit, Ref LK043.L.OPT.A, IVD. The firm name on the label is The Binding…
The Binding Site Group, Ltd.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: The Binding Site Group, Ltd. · 8 Calthorpe Road · Birmingham, N/A · N/A
Is this the product you have?
- Binding Site Optilite Beta-2-Microglobulin Urine Kit, Ref LK043.L.OPT.A, IVD. The firm name on the label is The Binding Site Group Ltd., Birmingham, U.K. The Optilite Beta-2-Microglobulin (¿2M) Urine kit is intended for the quantitative in-vitro measurement of beta-2-microglobulin in urine using the Binding Site Optilite analyser to aid in the diagnosis of active rheumatoid arthritis and kidney disease. This test should be used in conjunction with other laboratory and clinical findings.
- Lot codes and dates
Lot 407460UDI 05051700017688
- Amount recalled
- 20 kits distributed to the U.S.
Where it was sold
Why it was recalled
The product is not meeting the antigen excess protection claim of up to 760mg/L, as stated in the Instructions for Use.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new The Binding Site Group, Ltd. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.