Reported Apr 1, 2015 · Terminated

da Vinci Xi Surgical System, IS4000; Assy, Distal SUJ, Inner, IS4000 Material number: 380662-20 and 380662-21. The…

Intuitive Surgical, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Intuitive Surgical, Inc. · 1266 Kifer Rd Bldg 100 · Sunnyvale, CA · 94086-5304

Is this the product you have?

  • da Vinci Xi Surgical System, IS4000; Assy, Distal SUJ, Inner
  • IS4000 Material number380662-20 and 380662-21. The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endocopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endscopic retractors, electrocautery and accessories for edoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximations, ligation, electrocautery, suturing and delivery and placement of microware and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic labaroscopic surgical procedures, general thorascopic surgical procedures and thorascopically-assisted cardiotomy procedures
Lot codes and dates
  • Material number3680662-20:
  • Serial numbers
    • 1311140273
    • 1311140370
    • 1311140385
    • 1311140148
    • 1311140099
    • 1311140152
    • 1311140081
    • 1311140473
    • 1311140561
    • 1311140409
  • Material number380662-21:
  • Serial numbers
    • 1311140225
    • 1311140437
    • 1311140476
    • 1311140336
    • 1311140472
    • 1311140438
    • 1311140353
    • 1311140195
    • 1311140545
    • 1311140098 1311140189
    • 1311140465
    • 1311140394
    • 1311140366
    • 1311140008
    • 1311140115 1311140014
    • 1311140371
    • 1311140238
    • 1311140306
    • 1311140332
    • 1311140404 1311140054
    • 1311140188
    • 1311140324
    • 1311140453
    • 1311140408
    • 1311140237 1311140220
    • 1311140372
    • 1311140325
    • 1311140088
    • 1311140535
    • 1311140355
Amount recalled
167 total da Vinci Xi affected systems

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

  • Nationwide

Why it was recalled

Correction due to a fault that causes distal SUJ (DSUJ) failures that indicates a mismatch between the positions reported by the redundant position sensors on one of the patient cart arm joints.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Intuitive Surgical, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Sep 8, 2015 · Data as of Sep 19, 6:03 PM UTC