Reported Jul 6, 2022 · Ongoing
Yukon Straight Rod 4.0 x120mm; Cat# 7601-540120. Used to provide immobilization and stabilization of spinal segments.
K2M, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: K2M, Inc. · 600 Hope Pkwy Se · Leesburg, VA · 20175-4428
Is this the product you have?
- Yukon Straight Rod 4.0 x120mm; Cat# 7601-540120. Used to provide immobilization and stabilization of spinal segments.
- Lot codes and dates
- UDI-DI10888857326804 Lot PCMW
- Amount recalled
- 53 devices (19 US; 34 OUS)
Where it was sold
Why it was recalled
Some of the rods in this lot were observed to have discoloration, later confirmed to be indication of trace amounts of inorganic phosphorous compound. This may produce a mild irritating or inflammatory response; adverse local tissue reactions may occur related to cytotoxicity with the nonconforming rod and surrounding tissue.
This product could cause temporary or reversible health problems — stop using it.