Reported Feb 11, 2026 · Ongoing

Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501, (2) M1502, (3) M1504, (4) M15050,…

Etac A/s

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: ETAC A/S · Parallelvej 3 · Gedved, N/A · N/A

Is this the product you have?

  • Brand NameMolift
  • Product NameMolift
  • Mover 300 Model/Catalog Numbers
    • (1) M1501
    • (2) M1502
    • (3) M1504
    • (4) M15050
    • (5) M15151
    • (6) M15001
    • (7) M15002
  • Product DescriptionThe mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength
  • ComponentThe deformation occurs between the bolt and the fork in the joint connecting the lifting bar and the arm on the Molift Mover
Lot codes and dates

21 of 21 products

Products covered by this recall
Model / numberProduct as published
M1501(1) Model/
UDI-DI: TBD
Serial Numbers: 601005, 601002, 611506, 621101, 621106, 621104, 621105, 621102, 611505, 611504, 611503, 611502, 611501, 601008, 611507, 601007, 601006, 601001, 621103, 601003, 601004
M1502(2) Model/
UDI-DI: TBD
Serial Numbers: 611409, 611408, 611407, 611406, 611405, 611404, 611403, 611402, 611401, 611412, 611411, 611410
M1504(3) Model/
UDI-DI: TBD
Serial Numbers: 600902, 600901, 611303, 611302, 611301
M15050(4) Model/
UDI-DI: TBD
Serial Numbers: 586003, 586001, 585905, 585903, 585902, 585901, 598605, 586101, 585904, 603602, 591401, 586002, 586006, 586005, 586004, 598608, 598607, 585906, 591601, 598606, 598604, 598602, 598601, 596501, 603601, 598609, 598603
M15151(5) Model/
UDI-DI: TBD
Serial Numbers: 599703, 599702, 599701
M15001(6) Model/
UDI-DI: TBD
Serial Numbers: 584401
M15002(7) Model/
UDI-DI: TBD
Amount recalled
71

This recall covers 4 products

The agency filed one notice per product. You are reading one of them; the other 3 are below.

Where it was sold

  • Nationwide
  • PA

Why it was recalled

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new packaged foods recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:41 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.