Reported Apr 9, 2014 · Terminated
TDWorkstation version 11.01.A or higher. TDWorkstation Evolution. TDWorkstation receives patient results from a…
Technidata S.A.
Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Technidata S.A. · 387, Avenue Jean Kuntzmannn · Montbonnot St Martin, N/A · N/A
Is this the product you have?
- TDWorkstation version 11.01.A or higher. TDWorkstation Evolution. TDWorkstation receives patient results from a connected instrument and sends them to a Laboratory Information System (LIS).
- Lot codes and dates
- version 11.01.A or higher
- Amount recalled
- 85 units total (1 unit in US)
Where it was sold
Why it was recalled
If a conflict occurs between a Demography Query and a Result Transmission by the instrument, the TDWorkstation could send an incorrect patient demography to an instrument.
This product is unlikely to cause harm, but it breaks a safety or labelling rule.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Technidata S.A. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.