Reported Mar 27, 2024 · Ongoing

Preat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967

Preat Corp.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Preat Corp · 2625 Skyway Dr Ste B · Santa Maria, CA · 93455-1405

Is this the product you have?

  • Preat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967
Lot codes and dates
  • Model Number9000967
  • UDI-DI code00842092165867
  • Lot Numbers
    • 235650
    • 234294
    • 233667
    • 233885
    • 233439
    • and 231489
Amount recalled
561 units

Where it was sold

Why it was recalled

Due to a manufacturing issue, the screw seat location on the abutment product may be too high, causing mating screw to have a shorter engagement length into the associated implant.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.