Reported Apr 13, 2016 · Terminated

The THERAKOS CELLEX Photopheresis System is a photoimmune therapy. Catalog CLXUSA. Manufacturing lot D342 UDI:…

Therakos Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Therakos Inc · 10 N High St · West Chester, PA · 19380-3003

Is this the product you have?

  • The THERAKOS CELLEX Photopheresis System is a photoimmune therapy. Catalog CLXUSA
  • Manufacturing lot D342 UDI20705030200003 The system designed to separate the patient's blood into various fractions and collect the white blood cell (WBC) fraction. The WBC fraction is treated with a photoactive drug UVADEX¿ (methoxsalen) Sterile Solution. The UVADEX is photo-activated by UVA light and results in the WBCs undergoing apoptosis
Lot codes and dates
  • Mfr lot #D342
Amount recalled
729 units

Where it was sold

Why it was recalled

The Mallinckrodt (Therakos) post-market monitoring process detected an increased trend in tubing leaks within the pump tubing organizer, associated with a portion of Kits from Lot D 342.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Therakos Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Oct 14, 2016 · Data as of Sep 19, 5:20 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.